CDN 2 lines | CWG 1526
IEC / EN 61000-4-5
With the help of the coupling network of type CWG 1526, EMC tests (immunity tests) can be carried out on electrical consumers. These tests are based on IEC 61000-4-5 (surge test for unscreened, unbalanced connection lines).
The interference signals of the CWG 1500 / CWG 2500 surge generator are superimposed on the connecting lines of the instrument under test. The coupling switch can be used to select the interference paths (D1/GND, D2/GND or D1/D2).
- Nominal voltage AC: max. 240 V, 50 / 60 Hz
- Nominal voltage DC: max. 120 V
- Rated current IN at TU= 40 °C:
CWG 1526-4: 2 x 4 A
CWG 1526-10: 2 x 10 A - Series choke: 2 x 20 mH
- Coupling capacitance C: 0.5 µF
- Serial resistance R: 40 Ohm
- Access method: D1 - D2; D1 - earth; D2 - earth
- Maximum pulse voltage: 1.2/50 µs 4400 volts
- High voltage (HV) - Input: Fischer HV socket D105A039
- Input / Output coupling network: Lab jacks
- Electronics supply: 100 - 240 V AC 47 - 63 Hz, 50 VA
- Earth connection: additionally via socket on the front and rear
- Operating temperature: 0 to 40 °C
- Weight: CWG 1526-4: 4.8 kg CWG 1526-10: 10.4 kg
- Dimensions (H x W x D): 150 x 225 x 360 mm (3 HE; 42 TE)
The CWG 1526 is suitable for operation with the CWG 2500 / 1500 surge generators from Schlöder.
HV cable with 0.85 m length for connection to the Surge Generators: CWG 1500 / CWG 2500