Coupling network CN 19 U with current transformer CN 19 I

IEC / EN 61000-4-19

The coupling network CN 19 U is suitable for differential voltage tests according to IEC / EC 61000-4-19. It is designed for tests in the frequency range from 2 kHz to 150 kHz.

The current transformer CN 19 I is used to supplement tests according to the IEC / EC 61000-4-19 standard and is also suitable for a frequency range from 2 kHz to 150 kHz.

Kontakt / Contact
We will gladly advise you personally!

CN 19 U

  • Frequency range: 2 kHz – 150 kHz
  • Level: 0...25 Vrms cont
  • Source impedance: 10 Ω ±3 Ω max. (±1 Ω typical)
  • EUT voltage: 500 V AC (50/60 Hz), 700 V DC
  • Max. EUT current: 16 A cont. AC/DC
  • Voltage drop mains EUT: 13 V typical (16 A/50 Hz)
  • Amplifier input: XLR-female 
  • Mains/EUT: 4 mm safety sockets (black/blue)
  • High voltage test amplifier - EUT: 1.9 kV DC (10 sec)
  • Dimensions (H x W x D): 15.7 x 48.2 x 54 cm
  • Weight: 11.2 kg
  • Temperature range: 5...40 °C
     

CN 19 I: current transformer

  • Frequency range: 2 kHz – 150 kHz
  • Level: 0...10 Arms cont.
  • Source impedance: 1 Ω ±0.3 Ω max. (±0.1 Ω typisch)
  • Four-wire shunt: 0.1 Ω ±1 mΩ max.
  • EUT voltage: 500 V AC (50/60 Hz), 700 V DC
  • Amplifier input: XLR-female
  • EUT current output: 4 mm safety sockets (black/blue)
  • Shunt output: 4 mm 4 mm safety sockets (green)
  • Power supply: external 24 V/1 A AC/DC-power supply unit
  • High voltage test amplifier - current output: 1.9 kV DC (10 sec)
Accessories